Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J B-5
This section lists the technical details for all parameters and functions in
the Micro-Tech’s “Main Menu 1” that have not been covered elsewhere
in the manual.
For detailed step-by-step instructions about doing a zero calibration on
your scale, see the “Doing a Zero Calibration” section on page 3-5.
The zero number is a digital representation of the tare load on the scale.
The tare load consists of the scale carriage, idler(s), and the section of
belt within the weigh area. The zero number displays the load-cell signal
(mVDC) in digital form.
Here is the Zero Calibration menu table.
Parameter
Selection
Description
ZERO CAL
START, EXIT, MANUAL
Run belt empty, then press
START
AUTO ZEROING
ABORT
Time remaining
Rate
Tot
AUTOZERO COMPLETE
YES, NO, ADV
XXXX = Error%, Total
XXXX
Change Zero?
ZERO # CHANGED
RUN, MENU, ADV
XXXX = Old zero #,
Rate, Total, Error
New zero #
XXXX
MANUAL ZERO EDIT, EXIT, ADV (Password: Operator)
Rate
Zero # XXXXX
MANUAL ZERO
EXIT, ADV
AZ Dev - Shows the
difference between the new
zero number and the Zero
AZ Dev
Main Menu 1
Zero Calibration