B-6 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
Parameter
Selection
Description
Total AZ %
Reference number.
Total AZ % - Shows the
total deviation of the new
zero number from the Zero
Reference value.
MANUAL ZERO
EXIT, ADV
Note: This screen is only
available when Auto zero
tracking is turned on under
MAIN MENU 2, CALIB
DATA.
AZT # - Shows the
adjustment to the Zero
Number.
AZT % - Shows the total
deviation of the AZT
correction factor from the
Zero Reference value.
AZT #
AZT %
The Micro-Tech has three calibration methods, as follows.
The electronic R-Cal method. (See page 3-7 for details.)
Static-weights method. (See below.)
Test-chain method. (See below.)
To change the calibration method, do the following.
1. Make sure the Micro-Tech Run screen is currently being displayed.
2. Press the Menu button twice to bring up “Main Menu 2.”
Span Calibration
Changing the
Calibration Method
BATCH
0.0 Tons
0.0 Tph
READY
ALARM
CALIB
TOTALS