Operation
Calibrating the Micro-Tech
3-10 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
4. Make sure the belt is running empty at maximum speed, then press
the Start button. The count-down screen appears. (The data in your
screen will, of course, be different.)
NOTE. The number of seconds shown in “Time remaining” is
calculated based on the current speed-sensor pulse frequency, and
estimates the time remaining for a complete test.
5. The calibration time (in seconds) that you established during the
Micro-Tech cold-start procedure, will start counting down. When the
counter reaches zero, the calibration is complete and the change-span
screen appears.
Time remaining 375
Rate 150 Tph
Tot 0.000 Tons
Change span?
Error 0.01%
NO YES ADV