Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J B-61
Here is the Linearization menu table.
Note: No linearization is done until you perform
an Auto Multipoint material calibration or
manually enter the linearization factors.
(Password: Service.)
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 20%.
Enter a Factor and a Load.
If you enter 0.000000 for the factor, the rate is
not corrected in that portion of the range. A
factor lower than 1.000 reduces the span, while
a factor larger than 1.000 increases the span.
The ACQ key automatically acquires the Load
at the present flow rate. (Password: Service.)
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 40%.
"
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 60%.
"
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 80%.
"
LINEARIZ. #5
Load ______
Fact. ________
0.00 Lb/Ft
1.000000 (0-1.5)
This scroll is for flow rate under 100%.
"