EasyManua.ls Logo

Thermo Scientific Micro-Tech 9101 - Page 181

Thermo Scientific Micro-Tech 9101
329 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J B-9
Note that test chains have the weight per foot stamped on the link at
the end of the chain. Press the Edit button and use the keypad to
enter the appropriate value.
Static-Weights Method
Add up the total weight you added to the scale, press the Edit button,
and use the keypad to enter the appropriate value.
BATCH
- CALIB DATA SCROLL 6 -
Total test weight on
scale 0.000 lbs
READY
ALARM
CALIB
EDIT

Table of Contents

Related product manuals