B-52 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
Alarm
Number
Description
Selection
Description
and the printer is
disconnected or paper is
empty.
#55
Communication error
ALARM, SHUT DOWN, AUTO
RES., NONE
Not present when Printer is
selected for Protocol port #1.
This alarm indicates a time
out or handshake error is
detected during a data
transfer on the COMM line.
#57 PROFIB comm. error ALARM, SHUT DOWN, AUTO
RES., NONE
This scroll is only available if
the optional Profibus board is
installed. The following two
conditions activate the alarm.
The Siemens SPC3
Controller installed on the
Profibus interface board does
not recognize any successful
data transfer within the
watchdog timer interval.
The received data contains
errors (value overlaps limits,
register number does not
exist, group number does not
exist).
#64
Load cell imbalance
NONE, ALARM, SHUT DOWN,
AUTO RES.
#66
Auto Zero Limit
ALARM, SHUT DOWN, AUTO
RES., NONE
#69 AZT step over limit ALARM, SHUT DOWN, AUTO
RES., NONE