Operation
Calibrating the Micro-Tech
Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J 3-7
3. Press the Zero Calibration button and the following screen appears.
4. Make sure the belt is running empty and at maximum speed, then
press the Start button. The count-down screen appears. (The data in
your screen will, of course, be different.)
NOTE. The number of seconds shown in “Time remaining” is
calculated based on the current speed-sensor pulse frequency, and
estimates the time remaining for a complete test.
5. The calibration time (in seconds) that you established during the
Micro-Tech cold-start procedure, will start counting down. When the
counter reaches zero, the calibration is complete and the change-zero
screen appears.
BATCH
Run the belt empty, then
press START
READY ALARM CALIB
Time remaining 375
Rate 0.00 Tph
Tot 0.000 Tons