Operation
Calibrating the Micro-Tech
3-12 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
6. This is important! Press the No button (that is, do not change the
span) and the following screen appears. (Your numbers will, of
course, be different.)
7. Press the Run soft key in the display to return the Micro-Tech to the
Run mode.
8. Run several span calibrations to assess the repeatability of the
readings.
BATCH
SPAN # UNCHANGED
New span # 199980
Old span # 199980
READY
ALARM
CALIB
MENU
RUN
ADV