Table 39 List of Configuration Mode Commands
Command Description
?
Displays a list of configuration commands.
BAUD
Changes or displays serial port settings.
CLEARERR
Resets error counters.
CLOSE
Switches the serial port to measurement mode.
ERRORS
Fetches the error codes and counter information from WMT700.
G
Displays either all or specified parameters.
H
Displays a list of data messages and available values for measurement unit, profile, baud
rate, interface, and analog output mode.
MEAS
Starts wind measurement based on the user-configurable averaging time. WMT700 does
not send data messages automatically.
POLL
Tests data polling.
RESET
Resets WMT700.
S
Changes selected parameters or defines new data messages.
START
Starts continuous measurement.
STOP
Stops continuous measurement.
VERSION
Displays the software version.
WIND_GET
Fetches wind calibration information.
More Information
‣
ERRORS — Get Error Codes and Counts (page 106)
‣
Ordering Options (page 18)
‣
Restoring Serial Port Settings (page 170)
5.4
Parameter Handling Commands
All the commands in the configuration mode are in the following form:
>CMD x,y<enter>
>
Configuration mode prompt
CMD
Command
Space
x
Any parameter in Table 73 (page 187)
WMT700 User Guide M211095EN-H
102