EasyManua.ls Logo

Vaisala WINDCAP WMT700 Series - 5.24.1 WS425 A;B ASCII Commands; 5.24.1.1 I - Identify Sensor

Vaisala WINDCAP WMT700 Series
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<CR>
Carriage return code, ASCII 0DH
<LF>
Line feed code, ASCII 0AH
To use WMT700 with the NMEA Extended profile, either set the autoSend parameter to
0 to enable polling or define a fixed output interval with the autoInt parameter. If you
are using automatic messages, the value for the autoSend parameter must be set to 15.
Table 52 Checksum Table
ID Character
<id>
Checksum
<chk>
Polling String
A 72
$WIPAQ,*72<CR><LF>
B 71
$WIPBQ,*71<CR><LF>
C 70
$WIPCQ,*70<CR><LF>
D 77
$WIPDQ,*77<CR><LF>
E 76
$WIPEQ,*76<CR><LF>
F 75
$WIPFQ,*75<CR><LF>
G 74
$WIPGQ,*74<CR><LF>
H 7B
$WIPHQ,*7B<CR><LF>
I 7A
$WIPIQ,*7A<CR><LF>
J 79
$WIPJQ,*79<CR><LF>
K 78
$WIPKQ,*78<CR><LF>
L 7F
$WIPLQ,*7F<CR><LF>
M 7E
$WIPMQ,*7E<CR><LF>
N 7D
$WIPNQ,*7D<CR><LF>
O 7C
$WIPOQ,*7C<CR><LF>
5.23.2 WS425 A/B NMEA Extended Data Message
WS425 A/B NMEA Extended data message is as follows:
$P<id>MWV,<dir>,<ref>,<spd>,<uni>,<sta>*<chk><CR><LF>
$P
Fixed text
<id>
Data ID; A … Z
WMT700 User Guide M211095EN-H
146

Table of Contents

Other manuals for Vaisala WINDCAP WMT700 Series