• WMT700 checks the status of the internal operational software at each reset by
verifying the CRC checksum of the software. The software check is not made
periodically; it is only made at each reset. If the checksum is incorrect, WMT700 does
not start up. If the checksum is correct, WMT700 continues by calculating the CRC
checksum of the parameter memory. In case the parameter memory is corrupted, the
sensor replies with the text Parameters crc error upon startup. In case the
parameter memory checksum is correct, WMT700 replies with the text ready as
shown in the example below.
>$WMT700 v<version number>
ready.
• To check the status of the wind sensor, include the status code item \er in the data
message. You need to convert the status code to a binary number to verify the status.
An error code does not indicate that the wind data is invalid.
• To check that the wind data is valid, include the \va item in the data message.
• To check other diagnostics-related data, include the required items in the data
message.
More Information
‣
Status Flags (page 113)
‣
Items for Data Messages (page 110)
‣
Command Set for WMT700 (page 181)
5.11
Measurement Mode Commands
The available commands in the measurement mode depend on the selected profile.
For a list of serial commands available for all the communication
profiles supported by
WMT700, see A. Command Set for WMT700 (page 181).
More Information
‣
Command Set for WMT700 (page 181)
‣
Operating WMT700 with WS425 and SDI-12 Profiles (page 133)
5.11.1
WMT700 Profile Commands
When you select the WMT700 profile, you can operate WMT700 with these commands. In
the command descriptions, <CR> is an ASCII carriage return control character and <LF> is
an ASCII line feed control character. You can send commands at any speed. The command is
executed when end-of-line character is received.
Chapter 5 – Operation
125