Figure 24 WMT700 North Arrow
Do not remove the instrument or sensor from the mounting kit during alignment.
1. Remove the transportation damper that protects the array and store it for future use.
2. To align the sensor, use a compass or other similar method to rotate the sensor so that
the North arrow points North.
3. Tighten the bolts of the mounting kit. Tightening torque 5 Nm.
4. If you cannot align WMT700, measure the deviation angle from North and configure
the wind direction oset error.
Connect the cable to the data acquisition system and power supply.