Character Description
15 Checksum (most significant digit). For more information, see the note below.
16 Checksum (least significant digit)
17 03H (<ETX>, end of transmission)
18 CR (carriage return)
19 LF (line feed)
The checksum is calculated from the characters from position 2 to 14. The accumulator
initializes at 0 with the addition of the byte value. The checksum has a range of 0H … FFH.
Example
W5 W5P1200013.2TDE
Interpretation of the example message:
02H (<STX>, start of transmission)
Sensor ID: W
Averaging in seconds: 5
Sensor status: P = pass
Wind direction: 120
Wind speed: 0013
. (dot character)
Wind speed unit: T = meters per second
Checksum (most
significant digit): D
Checksum (least significant digit): E
03H (<ETX>, end of transmission)
Missing Readings
If data is missing due to a measurement problem, the WS425 A/B ASCII message reports
999.9 for wind speed.
5.25
 WS425 A/B WAT11 Profile
The following table lists the configurable parameters and their allowed and default values
for the WS425 A/B WAT11 profile.
WMT700 User Guide M211095EN-H
150

Table of Contents

Other manuals for Vaisala WINDCAP WMT700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala WINDCAP WMT700 Series and is the answer not in the manual?

Vaisala WINDCAP WMT700 Series Specifications

General IconGeneral
BrandVaisala
ModelWINDCAP WMT700 Series
CategoryAccessories
LanguageEnglish

Related product manuals