EasyManua.ls Logo

WAGNER Titanus Micro-Sens - Page 29

WAGNER Titanus Micro-Sens
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TITANUS
MICRO·SENS
®
Technical Description
06/13
29
alarm point via a contact. The monitoring windows can be adjusted to the
environmental conditions.
The principle of the airflow sensor signalling process can be seen in the fol-
lowing figure.
Figure 9: Example signal pattern in the airflow sensor during faults
The detection unit is monitored for dirt and signal fault. Any fault which oc-
curs is displayed at the TITANUS
MICRO·SENS
®
and can be forwarded to
the FAS via a contact.
Airflow adjustment on the TITANUS
MICRO·SENS
®
is fully automatic when
the detection unit is inserted in the device base, if previously the Jumper X4
had been changed. This plug & play reduces the time needed for commis-
sioning to a minimum. It is also possible, however, to adjust the airflow using
the DIAG 3 diagnosis tool. This means the initialisation phase can be carried
out in an air pressure-dependent or air pressure-independent manner.
A fault message is reset via the connected FAS. If whilst the TITANUS
MI-
CRO·SENS
®
is operating an FAS alarm and fault [messages] have to be re-
set at the same time as the control line, then as an option a reset board can
be inserted in a separate housing. The reset board can only be placed in a
separate housing and if the idle current on the line is between 5 mA and 50
mA. The line must be switched to dead for resetting. The effect is that when
there is any short-term switching off of the line voltage, the alarm and fault
messages on the TITANUS
MICRO·SENS
®
are automatically reset.
Device monitoring
A
irflow adjustment
Resetting through FAS

Table of Contents

Related product manuals