3-10
IM PX8000-02EN
3.10 Triggering on an Edge On A Trigger (Enhanced)
This section explains the following settings for triggering on an Edge On A trigger.
Trigger source
State condition
State condition achievement condition
Edge detection condition
Trigger condition
Features Guide: “Edge On A Trigger (Enhanced)”
ENHANCED Edge On A Trigger Menu
Press SIMPLE/ENHANCED and then the Setting soft key to select Enhanced. The following menu
appears.
Set Setting to Enhanced.
Set the trigger hold off. ► section 3.3
Set Type to Edge On A.
Set the state and trigger conditions.
Setting the State and Trigger Conditions (Set Pattern)
Press the Set Pattern soft key to display the following menu.
Set the state condition achievement conditions
(True, False).
Set the state condition (H, L, or X (do not use as a trigger source)).
Set the edge detection condition ( , , —).
Set the trigger level.
Set the hysteresis ( , , ).