6-8
IM PX8000-02EN
6.5 Changing the Harmonics List Display (Option)
This section explains the following settings for the harmonics list display (Hrm List). These settings are
available on models with the harmonic measurement (/G5) option.
List number
Measurement function
Element and wiring unit
► Features Guide: “Single Harmonics and Dual Harmonics Lists (Hrm Single List/Hrm Dual
List; option)”
NUMERIC SETTING Menu
Use the DISPLAY MODE and DISPLAY SETTING keys as explained in section 5.1 to display the
NUMERIC SETTING menu.
Set the display format to Hrm Single List or Hrm Dual List.
Select the list number that you want to set (1, 2).
Function, element, and wiring unit settings that you make for list number 2 are also
reflected in the right column of the harmonic order data of the dual harmonics list.
Set the measurement function (U, I, P, S, Q, λ, Φ, ΦU, ΦI, Z, Rs, Xs, Rp, Xp).
Set the element and wiring unit (Element 1-Element 4, ΣA, ΣB).
Hrm Single List Display Example
Measurement function side Harmonic order data side
Note
If every numeric data is displayed as no data (−−−−−−−), check the referenced sections under “Note” in
section 6.1.