08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
4 Use the keypad to enter the lot number.
5 Touch to confirm the entered lot number.
6 Enter the expiration date (use two digits and lead-
ing zero, if necessary) and touch to confirm
the expiration date you have entered.
Once you have finished updating the linearity test infor-
mation, you can use the next screen to select the lot
number you are currently editing as the current lot num-
ber.
7 Touch to confirm that you want this lot
number to be the lot number currently in use,
or
touch to store the entries without making the
lot number the current lot number.
8 Continue entering additional lot numbers,
or
touch to return to the Main Menu.
123
456
789
A-O0
Linearity Lot
12:48 pm
03/17/17
123
456
789
0
Linearity Expiration
03/31/17
12:48 pm
03/17/17
The current lot number is provided automatically for use
with subsequent tests.
Linearity Lots
D-311: Do you want to
make Linearity lot 777678
the 'current' lot?
Make 'Current'
*
12/31/17
777732
Lin.
01/15/17 777723
Lin.
Exp. Date Lot Num.
Type
12:48 pm
03/17/17

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing

Related product manuals