08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Starting a glucose control test After preparing the meter as described, you can proceed
to the steps directly related to control testing:
1 From the Main Menu screen touch Control Test.
In the Control Test screen, the levels available for the
control solution are displayed. To the right of the buttons,
the word Required identifies the level at which a glucose
control test must be run to remove QC Lockout.
2 Touch Level 1 (Lo) or Level 2 (Hi) to select the level
for the following test. In the example above, Level 2
(Hi) is marked.
Control Test
Level 2 (Hi)
Level 1 (Lo)
Required
12:48 pm
03/17/17
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
03/17/17

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing