74
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Starting a glucose control test After preparing the meter as described, you can proceed
to the steps directly related to control testing:
1 From the Main Menu screen touch Control Test.
In the Control Test screen, the levels available for the
control solution are displayed. To the right of the buttons,
the word Required identifies the level at which a glucose
control test must be run to remove QC Lockout.
2 Touch Level 1 (Lo) or Level 2 (Hi) to select the level
for the following test. In the example above, Level 2
(Hi) is marked.
Control Test
Level 2 (Hi)
Level 1 (Lo)
Required
12:48 pm
03/17/17
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
03/17/17