10.7 Service menu
Deleting patient information This feature allows you to delete all patient-related data
(patient lists and results). This may become necessary to
comply with data protection and privacy policies when
sending a meter back to the manufacturer or transferring
it to another facility.
To ensure deleted data is not transferred back through an
unwanted automatic synchronization with the DMS,
remove the battery pack immediately after deleting
patient data.
The option Service Menu with the feature “Delete
Patient Information” is not available in the preconfig-
ured default setting. The Service Menu can only be acti-
vated via DMS.
Service Menu
Delete Patient Information
12:15 pm
03/17/17
Admin Menu
Date / Time
Language Selection
Setup Menu
12:48 pm
Service Menu
03/17/17
Delete Patient Information
Service Menu 12:15 pm
QC
Test Display
Password
Test Ranges
Operator ID Patient ID
Unlock Delete Pat.
03/17/17
Confi rm Deletion
D-201: Delete Data:
Are you sure you want to
delete all patients and all
results from this meter?
This operation cannot be
undone.
Be aware that the function Delete Patient Information
irrevocably deletes all data once you touch to
confirm. The system will delete any patient results
regardless of whether they have been synchronized
with a data management system or not. Therefore,
always ensure that you have downloaded all relevant
patient results before you use this function.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing