Linearity Testing • 7
103
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
7 Linearity Testing
7.1 Information regarding linearity tests
Observe the applicable regulations and directives of the
responsible regulatory agencies when performing linearity
tests. For information about sources for products required
during linearity testing or for technical assistance,
contact Accu-Chek Customer Care.
Linearity tests can help you to check the function and
accuracy of the entire system over the full range of
specified values. Linearity samples should be treated in
exactly the same manner as previously described for
control solutions used in glucose control testing.
The term “Linearity” describes the ability of the system to
maintain a constant accuracy over the full range of spec-
ified values. If test results were plotted against reference
values as a curve over the full range of these values, the
ideal (high linearity) would be a straight line. Linearity is
the range of values from the lowest to the highest for
which an instrument is proven capable of giving accurate
results.
When using a data management system for configura-
tion, it is possible to partially or fully disable the func-
tions described in this chapter. In this case, the
respective buttons in Main Menu 2 do not appear. See
also Appendix A.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)