08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Linearity testing intervals The linearity of the system should be checked before it is
used the first time for patient testing. The intervals for
subsequent linearity tests are determined by the facility
that operates the system. Linearity testing can also be run
when you want to check the overall performance of the
system.
Information stored during
linearity testing
The following information is stored for every linearity test:
Test result
Lot number of the linearity solution
Level of linearity solution (L1 to L6)
Operator ID (if configured)
Lot number of the test strips
Time and date of test
Comments (if applicable)

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing