134
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
9.7 Display options and optional tests
Use this menu to select the unit of measure for test
results and enable or disable optional tests.
1 Touch the desired option to enable it:
■ Result Units
– mg/dL
– mmol/L
■ Enable Tests
– STAT: STAT tests may be performed. When this
option is enabled, you must enter the number of
STAT tests allowed (after confirming the settings).
– Linearity is displayed in the Main Menu 2 screen
– Proficiency is displayed in the Main Menu 2 screen
– Maintenance is displayed in the Main Menu 2
screen
2 Touch to save the settings. If the STAT option
is enabled, enter the number of STAT tests allowed,
or
touch to exit this menu without saving any
changes.
The display returns in both cases to the Setup Menu
(after entering the number of STAT tests allowed).
Test Disp. Options
mg/dL
mmol/L
Enable Tests:
Stat
Linearity
Profi ciency
Maintenance
Display Strip Limitations
Warning
Result Units
12:48 pm
03/17/17
123
456
789
0
STAT Test Allowed
8
12:48 pm
03/17/17