08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Isolation Room Workflow If configured, this feature allows you to reverse the order
of the steps “Entering or selecting the patient ID” and
“Confirming or selecting the test strip lot” to avoid con-
tamination issues. This workflow allows you to test a
patient in an isolation room without having to take the
test strip vial to the patient.
Test procedure for patients in an
isolation room
1 Confirm or select the test strip lot. For more details
see page 56.
2 Take the meter and the test strip to the isolation
room.
123
456
789
A-O0
Patient ID
12:48 pm
03/17/17
Patient Test
Patient 123456789
Use Strip L
ot
12:48 pm
123456?
03/17/17
Once a test strip has been removed from the vial, the
blood sample must be applied within 3 minutes.