08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Isolation Room Workflow If configured, this feature allows you to reverse the order
of the steps “Entering or selecting the patient ID” and
“Confirming or selecting the test strip lot” to avoid con-
tamination issues. This workflow allows you to test a
patient in an isolation room without having to take the
test strip vial to the patient.
Test procedure for patients in an
isolation room
1 Confirm or select the test strip lot. For more details
see page 56.
2 Take the meter and the test strip to the isolation
room.
123
456
789
A-O0
Patient ID
12:48 pm
03/17/17
Patient Test
Patient 123456789
Use Strip L
ot
12:48 pm
123456?
03/17/17
Once a test strip has been removed from the vial, the
blood sample must be applied within 3 minutes.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing