168
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Meter reset A meter reset should only be performed if all other reme-
dies have failed.
1 Place the meter on a level surface.
2 Press the On/Off button for at least 12 seconds and
release the button.
The meter powers off and on again.
The Roche logo is displayed.
If the Roche logo does not appear within 15 seconds,
place the meter in a base unit for a minimum of
15 minutes to recharge the battery.
The meter performs a system check. During the
system check, the Versions screen appears for a few
seconds, then disappears. The Versions screen
shows the version numbers of the meter compo-
nents. (The Versions screen shown here is for illus-
tration purposes only. Version numbers on your
meter may differ.)
A message appears informing you that the date and
time settings have been lost due to the meter reset.
3 Confirm the message Loss of Date/Time.
Before performing any further tests, you must now
either enter date/time manually and then synchronize
the meter, or contact your system administrator for
help.
To enter date/time manually and synchronize the meter,
follow the instructions on the next page.
Kernel:
RootFS:
App:
Boot:
Build:
Versions
04.01.00
04.01.00
02.12.00
02.04.02
02.09.02
Checking system...
Performing self-checks...
Performing self-checks...
W-1302: D/T settings lost
due to meter reset or loss
of power. Before
performing further tests:
Enter D/T manually and
synchronize meter. OR: Ask
System Admin. for help.
Loss of Date/Time