Linearity Testing • 7
105
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Linearity test kit The linearity test kit contains glucose solutions in six
levels (6 vials, 2.5 mL each). For additional information
about the contents and handling of the kit, refer to the
package insert.
Preparing to run a linearity test Aside from special preparations (see the following sec-
tion), a linearity test is run the same as a patient test.
Please check the following:
At least one code file for test strips must be stored
in the meter and match the lot number of the test
strips used (see Chapter 6).
The proper test strips must be available.
An operator ID must be entered (with password, if
applicable), if the meter is configured for login.