08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Shut down meter Shutting down the meter shuts down wireless communi-
cation and all other functionalities. Date and time, how-
ever, are maintained. Powering up the meter will take
slightly longer than from power-off (standby) mode.
To shut down the meter, press the On/Off button for
about 5 seconds and release the button as soon as the
Roche logo is displayed and the meter beeps. The screen
goes blank and the meter is shut down.
Use meter shutdown when you want to remove or replace
the battery pack (see page 122).
Automatic shutdown In case of a Download Lockout (see page 157) or if the
battery pack is critically low (see page 165) , the meter
willl shut down automatically.
Battery Pack The meter contains a rechargeable battery pack that
begins charging as soon as it is placed in an active base
unit (i.e., one connected to a power supply).
If you press the On/Off button for too long, a meter reset
will be triggered after about 12 seconds (see page 168)
and date and time will be lost.
NOTICE
Use only the specially designed battery pack provided by
Roche Diagnostics
. Using any other type of battery may
damage the system.