EasyManuals Logo

Accu-Chek Inform II User Manual

Accu-Chek Inform II
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
78
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Applying the control solution Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply control
solution.
1 Wait until the flashing drop appears in the display
before applying the control solution. The meter
beeps.
2 Apply a drop of glucose control solution to the
front edge of the test strip. Do not apply the con-
trol solution to the top of the strip.
The control solution is pulled into the test strip by
capillary action.
Once sufficient control solution has been detected, the
meter beeps and the measurement begins.
Control Test
Control L2 (Hi) 123456
Strip L
ot 545794
12:48 pm
03/17/17
When applying the control solution, position the meter
so that the test strip port is always higher or on the
same level as the control solution. This prevents any
excess solution from flowing down the strip and enter-
ing the meter.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing