Glucose Control Testing • 4
75
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Confirming or selecting the lot
number for control solutions
Once you have selected the level, you are asked to con-
firm or enter the lot number of the control solution. Com-
pare the number displayed by the meter to the number on
the label of the control solution.
1 Select the lot number as follows:
To read the lot number from the control solution
bottle via barcode scanner, press and release
. Follow the instructions for scanning IDs
(see page 45).
If you want to use the preselected number dis-
played by the meter, touch to confirm.
If you want to use a different number than the lot
number displayed, touch to open the keypad
and enter the number manually, or
2 Touch to confirm the selected test strip lot
number.
For additional information about storing lot numbers for
control solutions, see page 94.
The meter can be configured so that manual confirma-
tion is not necessary. In this setup, only the lot number
is displayed. Additional options are not available.
The meter can be configured so that lot numbers are
entered via barcode scanner only.
Control Test
Control L2 (Hi)
123456?
Use Control L
ot
12:48 pm
03/17/17
123
456
789
0
Control Lot
12:48 pm
03/17/17

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing