88
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Transferring code key
information to the meter
The following description assumes that the meter is pow-
ered on and the Main Menu is displayed.
1 Touch to open the Main Menu 2 screen.
2 Touch Strip Lots to open the related menu.
3 Touch Add if you want to add the information for a
new test strip lot from a new code key. The Add
Strip Lot screen opens.
4 Insert the new code key into the opening of the
code key reader. The LED starts flashing green and
signals that the reader is ready to transfer data.
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
03/17/17
Main Menu 2
Maintenance
Strip Lots
Profi ciency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.
12:48 pm
03/17/17
Strip Lots
Add
Strip
06/31/17 855732
* 05/21/17 845678
Strip 06/21/17 845723
Type Exp. Date Lot Num.
Strip
12:48 pm
03/17/17
The code key reader remains ready to send for a few
seconds after it has transmitted data. As a result, you
can perform the following procedure on multiple meters
without having to reinsert the code key.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing