Linearity Testing • 7
109
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Applying a linearity test sample Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply linearity
solution.
1 Wait until the flashing drop appears in the display
before applying the solution. The meter beeps.
2 Apply a drop of the linearity solution to the front
edge of the test strip. Do not apply the solution to
the top of the strip.
The linearity solution is pulled into the test strip
area by capillary action.
Once sufficient linearity solution has been detected, the
meter beeps and the measurement begins.
Linearity Test
Linearity 12345678 L1
Strip L
ot 545794
12:48 pm
03/17/17
When applying the linearity solution, position the meter
so that the test strip port is always higher or on the
same level as the linearity solution. This prevents any
excess solution from flowing down the strip and enter-
ing the meter.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing

Related product manuals