08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Shipping
If the meter has to be shipped or transported over
long distances, always remove the battery pack
from the meter. This eliminates the possibility of the
battery pack overheating due to a short circuit. It
also prevents deep discharge and other damage to
the battery pack or meter.
Only ship undamaged battery packs. Damaged bat-
tery packs must be disposed of locally. See page 24
for the risks associated with damaged battery
packs and disposal information.
Package the battery pack for shipping so that it
cannot move around in the packaging. Also
observe any other applicable national regulations.
When shipping via third parties (e.g., by air or
parcel service), work with the carrier to check
whether specific requirements need to be met in
relation to the lithium-ion battery packs on the
basis of national or international laws on hazardous
goods and, where applicable, if special packaging
and labeling requirements apply.
Observe the following safety information when shipping
the meter and battery pack. Failure to do so may result in
injury to persons or damage to the meter or battery pack.
For short distances - example between a facility’s sites
- users may transport the battery packs (either installed
in the meter or separately) by road without having to
meet further requirements.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing