08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Obtaining a blood sample Prepare the selected blood collection site and obtain
blood from the patient per facility policy.
Recommendations for the collection of capillary
blood
If no facility policy exists for obtaining capillary blood, the
patient's hands (or heel in the case of small children)
should be washed with warm water and soap, and then
dried thoroughly. If you are using alcohol wipes or other
disinfectants when obtaining capillary blood, the patient's
skin must be completely dry before you lance the site to
obtain blood.
Use an auto-disabling single-use lancing device for each
patient. The lancing device must be intended for use by
healthcare professionals in a multiple patient setting.
Follow the manufacturer's instructions for use.
We recommend obtaining the capillary blood sample
from the side of the fingertip as this part is the least
sensitive to pain.
WARNING
Potential risk of incorrect results due to residues on
skin
Traces of food on the fingers or fatty residues from
hand cremes or soap products may contaminate the
sample and lead to incorrect results. Wash the punc-
ture site thoroughly and rinse with plenty of water.
Residues of water or disinfectant on the skin can
dilute the drop of blood and lead to incorrect results.
After you have washed and disinfected the site,
ensure that the patient's skin is completely dry before
lancing the site to obtain a capillary blood sample.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing