EasyManuals Logo

Accu-Chek Inform II User Manual

Accu-Chek Inform II
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
138
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Shift: QC Lockout is activated per shift.
Time Of Day: QC Lockout is activated at the speci-
fied time of day. (Up to six times per day can be
configured.)
Rotating QC: QC tests require alternating use of L1
and L2 controls, instead of requiring both controls
when performing a QC test.
2 Touch to save the settings. If the Hours or Strip
Count option is enabled, enter the desired number,
or
touch to exit this menu without saving any
changes.
The display returns in both cases to the Setup Menu.

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing