Appendix • A
183
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Reportable range message Whether to display a
reportable range
message (1) or not (0)
1NY
Reportable range message text 0 – 100 characters “Out of
Reportable
Range”
NY
Strip limitations warning Configuration, whether to
display the strip limita-
tions warning (1) or not
(0).
0YY
Strip lot verification 0: display only
1: yes/no confirmation
2: list selection
3: scan only
1YY
Additional Patient Test: enables consecu-
tive tests on the same patient (creating a
patient test series)
0: disabled
1: enabled
0NY
Blood sample type selection 0: disabled
1: enabled
0NY
Isolation Room 0: disable (select patient ID
before strip lot selection)
1: enable (select strip lot
before patient ID selection)
0NY
Linearity lot verification 0: display only
1: yes/no confirmation
2: prompt for lot entry
1YY
Display
Contrast 0 - 15 7 Y Y
Formats and Language
Date format 1: MM/DD/YY
2: DD.MM.YY
1YY
Time format 1: 24 hours
2: 12 hours
2YY
Language setting * 1: German
3: French
4: Spanish
5: Italian
6: Dutch
7: Swedish
8: English (USA)
9: Danish
11: Portuguese
8YY
Glucose result units 0: mg/dL
1: mmol/L
0YY
* Other languages: For the availability of languages not listed above, contact Roche.
Subject/Attribute Range Default Device DMS