08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Result deletion
Result deletion algorithm. Automatic or
First in First out.
0: automatic
1: FIFO
1NY
Result retention time (number of days
since a result was measured before it can
be deleted automatically if result deletion
algorithm is set to “automatic” and no
more memory is available).
1 – 1,000 d 30 N Y
Result download required 0: disabled
1: enabled
0NY
Electronic communication
Download warning 0 – 999 h 0: disabled N Y
Download Lockout 0 – 999 h 0: disabled N Y
Maximum number of list items transferred
in one POCT1-A message
1 – 500 75 N Y
Application timeout (within this time the
application expects a response by the DMS
to any POCT1-A command)
30 – 3,600 s 60 N Y
WLAN settings (general)
WLAN enabled 0: disabled
1: enabled
0NY
Use DHCP 0: disabled
1: enabled
0NY
IP (the meter‘s static IP-address if no DHCP
is used)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Subnet Mask (subnet mask to be used by
the meter if static IP is used. Note: if DHCP
is used; the DHCP-subnet mask will be
used.)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Use DNS 0: disabled
1: enabled
0NY
DMS host (the host’s DNS name if DNS is
used)
0 – 60 characters “” N Y
DMS IP address (the host’s IP address if no
DNS is used)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
DMS port (the port number to request
open socket from the DMS. Roche recom-
mends use of port numbers 1024.)
0 – 65,535 0 N Y
DNS IP (IP address of the DNS-server if
DHCP is not used and DNS is configured)
0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Gate IP (static gateway IP address) 0.0.0.0 – 255.255.255.255 0.0.0.0 N Y
Subject/Attribute Range Default Device DMS

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing