08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
You can add comments to a control test result (as with
blood glucose tests) (see page 65).
If you do not wish to add a comment to the test result,
touch to continue to the next level of the control
test, if necessary, or to return to the Main Menu.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives.
For control tests, comments are only required if the
result is outside the specified range (whether this is
displayed as a value or as FAIL). If the result is within
the specified range (displayed as a value or as PASS),
comments are optional.
Test results are saved automatically if the meter is pow-
ered off or if it powers itself off after 10 minutes of inac-
tivity/no screen touches (see “Automatic power-off” on
page 22).

Table of Contents

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)