256 1100 Series Nano Pump Service Manual
6 Control Module Screens for the Nano Pump
Pressure Test Use the F3-key (Pressure Test) to perform a pressure test of the system.
Several steps like purging the system, setting up Isopropanol as solvent on
channel A or A2 (if solvent selection valve is installed) and blocking the
column outlet with a blank nut have to be performed before operating the
pressure test. For details use the i-key to achieve context sensitive help, follow
the instructions on the screen and refer to "Running the Normal Mode
Pressure Test" on page 69. For evaluating the pressure test refer to "Evaluating
the Results" on page 71.