1220 Infinity II LC System User Manual 23
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications Agilent 1220 Infinity II LC Autosampler
Table 5 Performance Specifications Agilent 1220 Infinity II LC Autosampler
Type Specification Comments
Injection range 0.1 – 100 µL in 0.1 µL increments Up to 1500 µL with multiple draw
(hardware modification required)
Precision <1 % RSD of peak areas from 1 µL to
5µL
<0.25 % RSD of peak areas from 5 µL
to 100 µL
Measured with caffeine
Replicate
injections
1 – 99 from one vial
Pressure range 0 – 60 MPa (0 – 600 bar, 0 – 8702 psi)
Sample viscosity
range
0.2 – 50 cp
Sample capacity 100 × 2 mL vials in 1 tray
40 × 2 mL vials in ½ tray
15 × 6 mL vials in ½ tray (Agilent
vials only)
Carry over Typically < 0.1 %, < 0.05 % with
external needle cleaning
Injection cycle
time
Typically 50 s depending on draw
speed and injection volume
Minimum sample
volume
1µL from 5µL sample in 100µL
microvial, or 1 µL from 10 µL sample
in 300 µL microvial