1220 Infinity II LC System User Manual 25
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications Agilent 1220 Infinity II LC VWD
Table 7 Performance Specifications Agilent 1220 Infinity II LC VWD
Type Specification Comments
Detection type Double-beam photometer
Light source Deuterium lamp
Maximum data
rate
80 Hz
Short term noise
(ASTM)
<± 0.25·10
-5
AU at 230 nm
See Specification Conditions below
Drift
< 1·10
-4
AU/h at 230 nm
See Specification Conditions below
Linearity > 2 AU (5 %) upper limit See Specification Conditions below
Wavelength
range
190 – 600 nm
Wavelength
accuracy
± 1 nm Self-calibration with deuterium lines,
verification with holmium oxide filter
Slit width 6.5 nm typical over whole
wavelength range
Flow cells Standard: 14 µL volume, 10 mm cell
path length and 40 bar (580 psi)
pressure maximum
High pressure: 14 µL volume, 10 mm
cell path length and 400 bar
(5800 psi) pressure maximum
Micro: 1 µL volume, 5 mm cell path
length and 40 bar (580 psi) pressure
maximum
Semi-micro: 5 µL volume, 6 mm cell
path length and 40 bar (580 psi)
pressure maximum
Can be repaired on component level

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals