Performance Specifications
Photometric noise < 0.0002 AU rms Sixty consecutive scans on air with 0.5 s integration time at 0 AU, 500
nm; 11-point moving average: using equation:
Noise(rms)=SQRT((SUM(X-x)^2)/n) where x are measured values, X is
a 11-point moving average, n is the number of points
Photometric stability < 0.001 AU/h Scan on air at 0 AU, 340 nm, after 1-hour warm up, measured over 1
hour, every 60 s, integration time 5 s; difference between maximum
and minimum values are compared to specification; at constant
ambient temperature
Baseline flatness < 0.001 AU rms Scan on air at 0 AU, 340 nm, 0.5 s integration time
Typical scan time 1.5 s Full range
Shortest scan time 0.1 s Full range
Time until next scan 0.1 s Full range, 0.1 s scan, at least 150 consecutive scans
*
EP stands for European Pharmacopoeia
**
Apparent absorbance is strongly affected by dissolved oxygen. According to ASTM, bubble pure nitrogen through liquid for several minutes
immediately before use. Use only recently distilled water (not demineralized water).