Agilent 8453 UV-visible Spectroscopy System Operator’s Manual 81
Using your Agilent 8453 UV-visible Spectroscopy System 4
Controlling your Sipper System
Controlling your Sipper System
A sipper system transfers your sample by means of a peristaltic pump into a
flow cell for the measurement. To control your sipper system through the
Agilent ChemStation software, you have to adjust your current sampling
system for sipper introduction.
In addition, due to the length of tubing, the dead volume of your flow cell and
the flow rate of your pump, you adjust your sipper system parameters. For
details see your Installing and Operating Your Sipper System manual.
1 Select the Sipper in the Instrument Panel selection box.
2 Click Setup on the Instrument Panel. Type the path length of your flow cell
in cm and click OK.
3 Click Setup again and access the Sipper Parameter dialog by clicking
Parameter. The parameters required can be determined using the Flow Test
task of your Verification and Diagnostics mode.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals