Understanding Your Agilent ChemStation 191
Automation
8
Explicit and Cyclic Calibrations Together
• The first calibration line is for the same level but averages the calibration
parameters. The interval entry specifies that the recalibration is done after
every three samples.
• The second entry replaces all the recalibration parameters, that is, a
complete recalibration is done. It has no recalibration interval.
Sequence Table
The sequence table comprises seven lines. The first line specifies the cyclic
recalibration sample. The second line specifies the explicit recalibration that
is done only once at the beginning of the sequence. The third through seventh
lines specify the samples to be analyzed.
The order of the entries into the sequence table is very important. All cyclic
recalibration vial entries specifying cyclic calibration must appear before the
sample entries or any explicit recalibration entries for the method.
SimpReg Analysis Order
This section describes the analysis order for the SimpReg method.
Table 2 1 SimpReg Analysis Order
Seq Line Inj No. Method Vial Operation
2 1 SimpReg 1 Simple calibration
1 2 SimpReg 1 Regular calibration
3 3 SimpReg 2 Sample analysis
3 4 SimpReg 3 Sample analysis
4 5 SimpReg 4 Sample analysis
5 6 SimpReg 1 Regular calibration
6 7 SimpReg 5 Sample analysis
7 8 SimpReg 6 Sample analysis