116 Understanding Your Agilent ChemStation
5 Quantification
Correction Factors
Correction Factors
The quantification calculations use four correction factors, the absolute
response factor, the multiplier, the dilution factor, and the sample amount.
These factors are used in the calibration procedures to compensate for
variations in detector response to different sample components,
concentrations, sample dilutions, sample amounts, and for converting units.
Absolute Response Factor
The absolute response factor for a sample component represents the amount
of the component divided by the measured area or height of the component’s
peak in the analysis of a calibration mixture. The absolute response factor,
which is used by each calibrated calculation procedure, corrects for detector
response to individual sample components.
Multiplier
The multiplier is used in each calculation formula to multiply the result for
each component. The multiplier may be used to convert units to express
amounts.
Dilution Factor
The dilution factor is a number by which all calculated results are multiplied
before the report is printed. You can use the dilution factor to change the scale
of the results or correct for changes in sample composition during pre-analysis
work. You can also use the dilution factor for any other purposes that require
the use of a constant factor.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals