80 Understanding Your Agilent ChemStation
4 Integration
Peak Recognition
Peak Recognition
The integrator uses several tools to recognize and characterize a peak:
peak width,
peak recognition filters,
bunching,
peak recognition algorithm,
peak apex algorithm, and
non-Gaussian calculations (for example tailing, merged peaks).
Peak Width
During integration, the peak width is calculated from the peak area and
height:
Width = Area/Height
or, if the inflection points are available, from the width between the inflection
points.
Figure 7 Peak width calculation