176 Understanding Your Agilent ChemStation
8 Automation
Data File Naming in a Sequence
Data File Naming in a Sequence
Data file naming in a sequence can be done in the following ways:
automatic,
manual, or
prefix/counter.
Automatic Data File Naming During a Sequence
Sample Vials
For example 017-0103.D
where:
The first three digits are the vial number, for example, 017.
The fourth digit in liquid chromatography and capillary electrophoresis is a
separation hyphen (-); in a gas chromatograph, this will be either (F) for
front or (B) for back.
The fifth and sixth digits are the sequence line that defines the method
used, for example, 01 for the first sequence line.
The seventh and eighth digits are the injection number for this vial by the
method, for example, 03 for the third injection.
Blank Runs
For example NV--0499.D
where:
NV stands for no vial.
- is a separation hyphen.
0499 is the 99
th
blank run of sequence line 4.