8 Biochemical Analysis Software for Agilent ChemStation
1 Installation and Configuration
Installing the Biochemical Analysis Software
The biochemical analysis software (G1117A) is an add-on module and is based
on the general purpose software (G1115A). For details of the installation
please see your Agilent 8453 UV-visible Spectroscopy System - Installation
Guide.
If you want to add the Biochemical Analysis software package please follow
the steps below.
1 Insert your Agilent UV-visible ChemStation CD-ROM.
2 Start setup.exe.
3 Select Modify and press Next.
4 Check G1117AA/G1123AA Biochemical Add-On and press Next.
5 Enter the license number for your Biochemical Analysis software and press
Add License.
6 Press Next and wait until the Maintenance Complete dialog is displayed
and press Finish.
7 After the installation is completed, run Installation Qualification, which is
available in the UV-Visible ChemStations menu.
8 If the Installation Qualification is successfully completed, the biochemical
analysis software is ready to use.