Understanding Your Agilent ChemStation 223
Using the ChemStation Reports
10
Other Report Style Parameters
Other Report Style Parameters
Summed Peaks Table
The peak sum table is provided for certain applications in the petrochemical
and pharmaceutical industries that can be performed more effectively with
the following features:
• Sum the areas of peaks that lie within a range specified by the user
• Sum the areas of a range of peaks and perform the calculations using a
single multiplier
• Sum the areas of all peaks having the same name
When the report is generated, the ChemStation uses the peak sum table to
produce a peak sum report that is printed after the standard report
calculations with the exception of Norm% which is replaced by the peak sum
report.
Report Layout for Uncalibrated Peaks
To change the report layout for uncalibrated peaks, choose one of the
following in the Specify Report dialog box.
• Separately to report the uncalibrated peaks in a separate table if sorting by
retention/migration time is selected, or separate tables if sorting by signal
is selected.
• With Calibrated Peaks to report the uncalibrated peaks together with the
calibrated peaks.
• Do Not Report to suppress the reporting of uncalibrated peaks.