ChemStation Workflow 7
ChemStation Data Structure
1
ChemStation B.02.01 and higher
ChemStation B.02.01 and higher
In order to strengthen the association between data files and methods, the
following new data organization scheme has been implemented with
ChemStation B.02.01 and higher. When used with the ChemStation, the
Agilent OpenLAB Enterprise Content Manager (ECM) also makes use of
the new data concept, since the complete data set (sequence/methods/data
files) can now be transferred (archived) to ECM as one entity.
Figure 1 Sequence Acquisition B.02.01 and higher
The methods in the folder Chem32\1\methods serve as master methods, i.e.
during acquisition and data analysis, they remain unchanged.
Similarly, the sequences in the folder Chem32\1\sequence serve as sequence
templates that can be used to rerun (but not reprocess) a sequence several
times.
HZfjZcXZ
IZbeaViZ
BVhiZg
BZi]dY
6fj^h^i^dc
HZfjZcXZ
XdciV^cZg