Biochemical Analysis Software for Agilent ChemStation 9
Installation and Configuration 1
Configuring your Agilent 89090A Peltier Temperature Controller
If you are planing to use the thermal denaturation mode of your biochemical
analysis software, you have to configure the Agilent 89090A Peltier
temperature controller.
1 Start the UV-Vis Configuration Editor from the UV-Visible ChemStations
Menu.
2 If no Agilent 8453 Instrument has been configured yet, add New Instrument
as described in the Agilent 8453 UV-visible Spectroscopy System
Operator’s Manual.
3 Select Instruments… from the Configure Menu.
4 Enter Instrument Name, choose the Initial Screen Window Size and click
OK.
5 From the UV-Vis Modules list select TemCo.
6 Set an GPIB address of 20.
7 Press Add to add the Peltier Temperature controller to your current
configuration.
8 Press OK to close the Device Configuration dialog box.
9 Use the Exit command from the File menu to quit the Configuration Editor.
10 Press Yes to save your configuration before exiting.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals