192 Understanding Your Agilent ChemStation
8 Automation
Cyclic Calibration Sequences with Bracketing
Cyclic Calibration Sequences with Bracketing
In a cyclic calibrated sequence with bracketing, the calibration table used to
calculate the unknown quantitative results is generated by averaging the
results of the current calibration with those of the previous calibration. This
new calibration table is a more accurate representation of the instrument
response at the time the sample was analyzed.
Example
Consider the following situation:
• The instrument response is drifting.
• Three injections of an identical two-component mixture are specified.
• Two injections are specified as calibration samples, and one is specified as
a sample.
• The first and third are calibration samples.
• The second injection is a sample.
To obtain a precise quantitative result for injection two (the sample), an
interpolation must be made between the two calibration samples; see the
figure. This process is known as bracketing.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals